![]() The mean differences between OsiriX and real measurements were less than 0.1 mm in both replicates, and maximum differences were less than 0.3 mm. All measurements were recorded by a principal (replicate 1, trial 1) and a secondary observer (replicate 2, trial 1) and were repeated once by each observer (trial 2). ![]() The lengths between tunnel orifices were measured using Mitutoyo Digimatic digital calipers to establish the gold standard, and with the OsiriX program in 3D multi-planar reformatting mode for comparison. 3D-CT scans were made of 14 frozen pig knees with five transosseous holes in the metaphyseal portion of femur. We evaluated the accuracy and reliability of length measurement on CT with OsiriX software. However, when CT data are processed and handled manually using image processing programs to yield 3D image and coordinate value, systematic and random errors should be validated. There is a growing interest in three-dimensional computed tomography (3D-CT) as a research tool for the study of bone, joint anatomy, and kinematics.